| Detailed Product Description |
 |
|
System configuration
|
|
|
F78-IIIB
|
F78-IIIC
|
|
High Voltage Generator
|
FS202-2
|
FS202-3
|
|
X-ray Tube Assembly
|
XD51-20.40/125
|
XD51-20.40/125
|
|
X-ray Tube stand Assembly
|
ZZ-3
|
ZZ-3
|
|
Diagnostic Table
|
ZC15X-1A
|
ZC15X-2
|
|
Radiology Table
|
SC3-2
|
SC3-2
|
|
Collimator
|
XS2-1, XS1-1
|
XS2-1, XS1-1
|
High Voltage Generator FS202
FS202 GENERATOR CONFIGURATION: FS202-2 (F78-IIIB) FS202-3 (F78-IIIC)

- FSK202 series can be used for conventional fluoroscopy, routine radiography, spot film radiography, and bucky radiography when it is used with diagnostic table ZC15X. In addition, FSK202 can also be equipped with a radiographic table SC3-2 for conducting linear tomography.
- Advanced functions of FS202-3 (computer controlled)
Improved performance and stability Self-diagnosis and error code display Auto-protection to the tube capacity Preset exposure parameters
FS202 Generator Technical Specification:
l FS202 Technical Specification |
|
l FSK202 CONTROL CONSOLE
|
|
Power
|
380V/220V AC
|
|
Fluoroscopy Tube Voltage
|
45~110kV
|
|
Fluoroscopy Tube Current
|
0.5~5mA
|
|
Time
|
0.02~5s
|
|
Allowed Internal Resistance
|
≤0.3Ω
|
|
Frequency
|
50Hz
|
|
FSB302-1 HIGH VOLTAGE TRANSFORMER
|
|
Power Supply
|
Single Phase
|
|
Power Frequency
|
50Hz
|
|
Wiring
|
Four-Tube Bridge Connection
|
|
Maximum Primary Voltage
|
375V
|
|
Maximum Secondary Voltage
|
125kV
|
|
Excitation Current
|
≤4.5A
|
|
Maximum Instantaneous Power
|
29.6kW/9.25kW
|
|
Maximum Continuous Power
|
5kW
|
|
Nominal Electric Power
|
29.6kW/7.4kW
|
|
Dimension
|
387mmX501mmX582mm
|
X-ray Tube XD51-20.40/125 Focus 2.0mm/1.0mm Rotating speed 2800 r/min 125KV Small focus 20KW / Large focus 40kW Max.80000HU
Tube stand assembly ZZ-3
| Tube stand assembly Specification |
|
Adjustable range between ceiling and floor
|
2650~3200mm
|
|
Sliding in longitudinal (along ceiling and floor)
|
2000mm
|
|
Sliding in transverse (along rotation arm)
|
900~1140mm
|
|
Up-down movement (along stand)
|
500~1800mm
|
ZC15X Diagnostic Table
|
ZC15X Diagnostic Table Specification
|
|
Film Size and Exposure Procedure
|
5" X 7"
|
|
8" X 10"
|
|
10" X 12"
|
|
11" X 14"
|
|
Range of Movement
|
≥720mm ( longitudinally )
|
|
≥250mm ( transversely )
|
|
300mm for pressure direction
|
|
Pattern of Movement
|
Driven electrically
|
|
Filtering Grid
|
Oscillating type; grid ratio: r=8; grid density: N=40L/cm
|
|
Focal distance: 70cm; Size: 14" X 14"
|
|
Tabletop moving
|
500mm at head end
|
|
ROTATION OF TABLE BODY
|
|
Range of Movement
|
+90° ~ 0° ~ -5°
|
|
Pattern of Movement
|
Driven electrically, at a speed of about 90°/30s
|
|
MOVEMENT OF TABLETOP
|
|
Range of Movement
|
Able to be extended 30cm at the front end
|
|
Pattern of Movement
|
Driven electrically, at a speed of about 30cm/s
|

SC3-2 Radiographic Table
|
SC3-2 Radiographic Table Specification
|
|
Tabletop Length
|
2000mm
|
|
Tabletop Width
|
730mm
|
|
Tabletop Height
|
690mm
|
|
Tabletop Movement (Longitudinal)
|
1200mm
|
|
Tabletop Movement (lateral)
|
200mm
|
|
Strut Movement Along Table (Longitudinal)
|
1200mm
|
|
Tabletop Locking
|
electromagnetically locking, brake pressure ≥100N
|
|
Tabletop Locking Control
|
footswitch (normally closed)
|
|
Bucky Vibration Mode
|
Oscillating type
|
|
BUCKY SPECIFICATION
|
|
Focus
|
fo=100
|
|
Grid Ratio
|
r=8
|
|
Grid Density
|
N=40
|
|
Grid Size
|
14" X 17" ( 356mm X 432mm )
|
|
Film Loaded in Bucky
|
Max 14" X 17"
|
|
Distance between Tabletop and Bucky Film
|
≤80mm
|
|
Longitudinal Travel of Bucky
|
500mm
|
|
Bucky Locking
|
electromagnetically locking, brake pressure ≥100N
|
|
Input Power
|
220V, 50Hz
|

Collimator: XS1-1(ZZ-3) & XS2-1(ZC15X)
Collimators Specification |
|
|
XS1-1
|
XS2-1
|
|
Mode of Adjustment
|
Manuel
|
Electrically-Driven
|
|
Maximum Tube Voltage
|
150kV
|
150kV
|
|
Radiation Field [Max SID = 65cm]
|
350mm X 350mm
|
350mm X 350mm
|
|
Radiation Field [Min SID = 100cm]
|
0
|
≤50mm X 50mm
|
|
Time Limit of the Light Source
|
≤60s
|
N/A
|
|
Inherent Filtration
|
2 (mmAl)
|
N/A
|
|
External Dimension (length X Width X Height)
|
265x212x223mm
|
212x212x223mm
|
|
Weight
|
6.7 kg
|
6.1 kg
|
Item (optional):
Vertical bucky stand: LS-3
LS-3 Specification |
|
Vertical movement
|
780mm~1850mm
|
|
Film size
|
from 20cm*25cm to 34cm*43cm
|
|
BUCKY SPECIFICATION
|
|
Focus
|
fo=150
|
|
Grid Ratio
|
N=40
|
|
Grid Density
|
R=10
|

TV system: F113-4
|
F113-4 Specification
|
|
CCD detector:
|
400,000 pixels
|
|
Scanning Ratio:
|
2.1 Interlaced
|
|
Number of Scanning Line:
|
625TCL
|
|
Aspect Ratio:
|
4:3 or 1:1
|
|
Resolution:
|
Horizontal Center≥800 TVL; Vertical Center≥400 TVL
|
|
Video Bandwidth
|
≥10MHz
|
|
Signal-to-Noise Ratio
|
≥48dB
|
|
Gray Scale
|
≥10
|
|