| HF50-R High Frequency 50kW Radiographic System
Technical Data Generator GFS501-1
H.F. inverter type 25kHz
Power output 50kW (500mA at 100kV) Exposure voltage 40kV to 150kV 1kV/step Exposure mA 25mA to 500mA by step Exposure time 2ms to 5s by step Exposure mAs 1mAs to 500mAs X-ray tube connection 0.6mm/1.2mm H.T. cable 75kV, 16 meter
X-ray tube Siemens RAY-12 Focal spot Large focus 1.2mm; Small focus 0.6mm Anode heat storage capacity 230kHU Rotation speed 3000 rpm
Radiographic table SC3-3 + Tube stand ZZ-5
Table top Length: 2100mm Width: 800mm Height: 650mm Longitudinal travel: 900mm Transverse travel: 240mm Grid density: N40 F0: 100cm Grid ratio: r10
Tube stand Vertical travel: 520~1760mm Longitudinal travel: 2380mm Self-rotation 0°~±90°can be fixed at 90° X-ray tube assembly rotating around arm: -120°~ +120°,can be fixed at any angle
Collimator Manual (with Light Indication)
Option
Ion Chamber Three field, for automatic expose control Vertical Bucky stand LS-3
 Vertical movement 780mm~1850mm Film size from 20cm*25cm to 34cm*43cm
Installation data Line voltage connection 3/N/PE~380V (±10%) at 50Hz Inner-resistance <0.3Ω Single phase, 220V, 50Hz
|