Home Products Services Articles About Us Contact Us
Medical Equipment  |  Dental  |  Disposable  |  Operating Room   |  Medical Suppliers  |  Anaesthetic  |  Home Care  |  Mobility
Products:
You are here: Home >> Products >> Medical Equipment>>X-ray
Medical Equipment

50kW X-ray System HF51-6

 

Detailed Product Description

SYSTEM CONFIGURATION:

  1. High Frequency Voltage Generator: GFS502-8
  2. X-ray Tube Assembly: Siemens Ray 12
  3. Remote Diagnostic Table: ZC25SY-2
  4. TV System
  5. Collimator: XS5-1

High Frequency Voltage Generator: GFS502 -8
GFS502-8 is a high frequency 50kW generator. It can support up to two X-ray tubes. The generator can be used in conventional radiography and Fluoroscopy.

 GFS502-8 Technical Specification:


GFS502-8 Technical Specification

Radiographic kVp Range

 40 ~ 150kV ±(5%+1kVp)

mA Range

 0.1mAs to 500mAs

Exposure Time

 1ms to 10s ±(1%+0.1ms)

Reproducibility (Coeff. Of Variation)

 0.5

Auto Exposure Control

 2 ion Chambers

TOMO

 Tomographic capability

Buckys

 2

Anatomical Programing

 yes

Fluoroscopic kVp Range

 40 ~ 120kV ±(5%+1kVp)

Power Supply

 380V, 3 Phases

 


X-ray Tubes: Siemens Ray-12

 

 

 

The Siemens Ray-12 X-ray tube assembly is using SV150/22/54s rotating anode tubes for all routine diagnostic examinations with conventional or digital radiographic and fluoroscopic workstations.
X-ray Tube Assembly Specification:


X-ray Tubes Assembly : Siemens Ray-12

 

RAY-12

Rated Voltage

150kV

Rated Focal Spot Value

0.6

Cathode Filament Max. A/Max. Volt.

4.6A/9.0V

Optical anode Angle

12°

Anode Heat Storage Capacity

2,30,000HU

Heat Storage of Tube Assembly

1,350,000HU

Leak Radiation

0.6mSv/h

Inherent Filtration

1.0 mmAl

Auxiliary Filter

1.5 mmAl

Total Filtration of Tube Assembly

2.5 mmAl

High Voltage Connection

3-pin

Weight (with flange construction)

17.5kg

Weight (without additional components)

18kg

Remote Diagnostic Table: ZC25SY-2

 

 

 

 

 

The ZC25SY-2 is a remote controlled diagnostic table for conventional or digital radiographic and fluoroscopic X-ray examinations. The table can tilt from 90° to -25°. The distance between the focuses and film is 1.1m or 1.5m.
ZC25SY-2 Diagnostic Table Specification:


ZC25SY-2 Diagnostic Table

Table Tilting Range

 +90° ~ -25° at 4.4°/s

Table Length

 2100mm

Table Width

 625mm

Movement of Spot Film device

 740mm±10mm( 0° ~ 90° ); 340mm±10mm( 0° ~ 25° )

FFD

 Changeable 110cm, 150cm

Transfer Movement of Table Top

 220mm±10mm

Collimator

 Manuel Auto

Grid

 r10, f120cm

FILM DIVISION

20cm X 25cm

 

25cm X 20cm

 

25cm X 30cm

 

30cm X 25cm

 

28cm X 35cm

 

35cm X 28cm

 

35cm X 35cm

 

35cm X 43cm

 

X-ray Intensifier TV System: F113-2

     The F113-2 self-adaptive X-ray imaging intensifier television system (hereinafter abbreviated as TV system) is exclusively designed for fluoroscopy real time monitoring. The TV system can be installed conveniently on an X-ray examination table. It offers several significant advantages including high definition, high sensitivity, low inertia, high signal-to-noise ratio and storage capacity for up to 8 images. High quality X-ray images for each part of human body can be obtained using this system.

F113-2 X-ray Intensifier TV System Specification:


F113-2 X-ray Intensifier TV System Specification

Power Supply

 AC 220V ± 22V, 50Hz ± 1Hz

Working Mode

 Continuous

Scanning Ratio

 2.1 Interlaced

Number of Scanning Line

 625TCL

Number of Frame per Second      

 25   

Vertical Frequency

 50Hz

Horizontal Freuqency

 15625Hz

Aspect Ratio

 4:3

Video Camera Resolution

 Horizontal Center600TVL ; Vertical Center400 TVL

Video Bandwidth

 10MHz

Signal-to-Noise Ratio

 48dB

Gray Scale

 8

Output Impedence

 75 ohms

Power Consumption

 2.4W

Video Input

 1V p-p ( positive ) 75 ohms

Video Output

 1V p-p ( positive ) 75 ohms

Channel Bandwidth

 8.7MHz ( equivalent to 700TVL )

IBS Low-Level

 0.5V

IBS High-Level

 4V ( DC )

DIGITAL SIGNAL PROCESSOR

Sampling Frequency

 20M

No. of Quantized Levels

 8 bits, level 256

Storage of Frames

 Up to 8 frames

Effective Frames

 1024 pixel X 576 lines X 8

Overall Power Consumption

 8W

MONITOR

Resolution

 Horizontal Center800TVL ; Vertical Center400 TVL

Video Bandwidth

 12MHz

Signal-to-Noise Ratio

 48dB

Gray Scale

 8

Adjustable Range of Raster

 Φ190±10mm

Video Input Signal

 1.0 Vp-p [ positive ] , 75 ohms

Video Input Impedance

 75 ohms, high impedance

Power Consumption

 35W


Collimator: XS5-1
    XS5-1 is multi-page and adjustable collimators for X-ray radiation field. A rectangular radiation field with a desired size can be produced by adjusting the collimator within a specified range. It is easy to install and to operate. Therefore, this unit can effectively reduce the damage to human bodies caused by X-ray leakage.

       Collimators Specification:


Collimators Specification

 

XS5-1

Mode of Adjustment

Manual / Auto

Maximum Tube Voltage

150kV

Radiation Field [Max SID = 65cm] 

350mm X 350mm

Radiation Field [Min SID = 100cm]

0

Time Limit of the Light Source

60s

Inherent Filtration

1.5 (mmAl)

External Dimension (length X Width X Height)

250x224x250mm

Weight

9.5 kg

Imaging System: DSI-1000 (optional)
Imaging System DSI-1000 Specification:


Imaging System Technical Specification

CCD CAMERA

Resolution

1024 X 1024 X 12bits (1,000,000 pixels)

Iris Calibration

Automatic

IMAGING INTENSIFIER

Fields

3 fields, 12”/9”/6”

Center Area Resolution

48 lp/cm

MONITOR

Monitor Size

17" Medical Monitor (2sets)

Digital Signal Input

SVGA

Screen Resolution

1280 X 1024

Scanning Frequency

73Hz No blink

Scanning Band Width

100MHz

 WORKSTATION

PC Console

Multi-CPU Intel Xeon 2.4GHz

512MB RAM

80 GB Harddisk, storing over 60,000 pictures

SOFTWARE

Operating System

Windows 2000 Professional

Imaging Processing Functions

Digital Fluoroscopy Acquisition 30 fps

Digital Spot Acquisition 1~15 fps

LIH

Edge Enhancement

Real-time Noise Reduction, Digital Subtract

Grey Level Adjustment

Automatic/Manuel Window Level

Automatic/Manuel Electric Cut

Image Rotation and Mirror Image

Zoom & Roam

Static high resolution display

4 / 16 frames display on same screen

Image measurement

Image notations

Patient management

 