| Detailed Product Description |
 |
SYSTEM CONFIGURATION: High Frequency Voltage Generator: GFS502-8 X-ray Tube Assembly: Siemens Ray 12(2sets) Remote Diagnostic Table: ZC25SY-2 Radiographic Table:SC4-4 TV System Collimator: XS5-1, XS1-2
High Frequency Voltage Generator: GFS502-8 GFS502-8 is a high frequency 50kW generator. It can support up to two X-ray tubes. The generator can be used in conventional radiography and Fluoroscopy.
GFS502-8 Technical Specification:
GFS502-8 Technical Specification |
|
Radiographic kVp Range
|
40 ~ 150kV ±(5%+1kVp)
|
|
mA Range
|
0.1mAs to 500mAs
|
|
Exposure Time
|
1ms to 10s ±(1%+0.1ms)
|
|
Reproducibility (Coeff. Of Variation)
|
≤0.5
|
|
Auto Exposure Control
|
2 ion Chambers capability
|
|
TOMO
|
Tomographic capability
|
|
Outputs
|
2
|
|
Anatomical Programing
|
yes
|
|
Fluoroscopic kVp Range
|
40 ~ 120kV ±(5%+1kVp)
|
|
Power Supply
|
380V, 3 Phases
|
X-ray Tubes: Siemens Ray-12

X-ray Tube Assembly Specification:
|
X-ray Tubes Assembly : Siemens Ray-12
|
|
Model Number
|
RAY-12
|
|
Rated Voltage
|
150kV
|
|
Rated Focal Spot Value
|
0.6 / 1.2
|
|
Optical anode Angle
|
12°
|
|
Rotation speed
|
3000 rpm
|
|
Anode Heat Storage Capacity
|
230,000HU
|
|
Heat Storage of Tube Assembly
|
1,350,000HU
|
|
Net Weight
|
17.5 kg
|

Remote Diagnostic Table: ZC25SY-2
The ZC25SY-2 is a remote controlled diagnostic table for conventional or digital radiographic and fluoroscopic X-ray examinations. The table can tilt from 90°to -25°. The distance between the focuses and film is adjustable ranging from 1.1m to 1.5m.
ZC25SY-2 Diagnostic Table Specification:
ZC25SY-2 Diagnostic Table |
|
Table Tilting Range
|
+90° ~ -25° at 4.4°/s
|
|
Table Length
|
2100mm
|
|
Table Width
|
625mm
|
|
Movement of Spot Film device
|
740mm±10mm( 0° ~ 90° ); 340mm±10mm( 0° ~ 25° )
|
|
FFD
|
Changeable 110cm, 150cm
|
|
Transfer Movement of Table Top
|
220mm±10mm
|
|
Collimator
|
Manuel Auto
|
|
Grid
|
r10, f120cm
|
|
FILM DIVISION
|
|
20cm X 25cm
|
|
|
25cm X 20cm
|
|
|
25cm X 30cm
|
|
|
30cm X 25cm
|
|
|
28cm X 35cm
|
|
|
35cm X 28cm
|
|
|
35cm X 35cm
|
|
|
35cm X 43cm
|
|
SC4-4 Radiographic Table

SC4-4 radiographic table, together with X-ray Tube Assembly, HV transformer assembly and console, can be used to perform conventional radiography, bucky radiography and oblique radiography. It can also be used together with vertical radiographic stand for chest filming.
SC4-4 Radiographic Table:
SC4-4 Radiographic Table Specification |
|
Tabletop Length
|
2000mm
|
|
Tabletop Width
|
740mm
|
|
Tabletop Height
|
658mm
|
|
Tabletop Movement (Longitudinal)
|
1200mm
|
|
Tabletop Movement (lateral)
|
200mm
|
|
Strut Movement Along Table (Longitudinal)
|
1200mm
|
|
DIS Travel while X-ray Tube Assembly Vertically Moving up and down
|
1100mm
|
|
Rotation Range of X-ray Tube Around Lateral Arm
|
+120° ~ -120°
|
|
Strut Rotation Range
|
+180° ~ -180°
|
|
Tabletop Locking
|
electromagnetically locking, brake pressure ≥100N
|
|
Tabletop Locking Control
|
footswitch (normally closed)
|
|
Bucky Virbration Mode
|
Fixed type
|
|
BUCKY SPECIFICATION
|
|
Focus
|
fo=100
|
|
Grid Ratio
|
r=10
|
|
Grid Density
|
N=40
|
|
Grid Size
|
14" X 17" ( 356mm X 432mm )
|
|
Film Loaded in Bucky
|
Max 14" X 17"
|
|
Distance between Tabletop and Bucky Film
|
≤60mm
|
|
Longitudinal Travel of Bucky
|
500mm
|
|
Bucky Locking
|
electromagnetically locking, brake pressure ≥100N
|
|
Input Power
|
220V, 50Hz
|
|
Optional Ion Chamber
|
Vacutec 70151
|
Collimator: XS5-1 & XS1-2
XS5-1 and XS1-2 are multi-leaf and adjustable collimators for X-ray radiation field. A rectangular radiation field with a desired size can be produced by adjusting the collimator within a specified range. It is easy to install and to operate. Therefore, this unit can effectively reduce the damage to human bodies caused by X-ray leakage.
Collimators Specification:
Collimators Specification |
|
|
XS1-2
|
XS5-1
|
|
Mode of Adjustment
|
Manuel
|
Manual / Auto
|
|
Maximum Tube Voltage
|
150kV
|
150kV
|
|
Radiation Field [Max SID = 65cm]
|
350mm X350mm
|
350mm X 350mm
|
|
Radiation Field [Min SID = 100cm]
|
0
|
0
|
|
Time Limit of the Light Source
|
≤60s
|
≤60s
|
|
Inherent Filtration
|
1.5 (mmAl)
|
1.5 (mmAl)
|
|
External Dimension (length X Width X Height)
|
250x224x250mm
|
250x224x250mm
|
|
Weight
|
9 kg
|
9.5 kg
|
AEC Ionization Chambers (optional)
 14" Monitor with Cart (optional)
Resolution: Horizontal Center≥800TVL ; Vertical Center≥400 TVL Video Bandwidth ≥12MHz Signal-to-Noise Ratio ≥48dB Gray Scale >8 Adjustable Range of Raster: Φ190±10mm Video Input Signal: >1.0 Vp-p [ positive ] , 75 ohms Video Input Impedance: 75 ohms, high impedance Power Consumption ≤35W
|